| Spinning
at 1200 RPM, air bubbles created from the agitation of the DI Water run down the
drain tube and do not dissipate by the time they wash past the resistivity probe,
causing a false reading.
To
alleviate this run rinse RPM within the limits set in the manual (300-600 RPM).
|